USSRPhoto.com

Soviet Photo Archive / 1926 No. 6

Microphotography

A. M. Belousov · how-to · original pages 176–177

The task of microphotography is the photographing of objects through a microscope — that is, in magnified form.

The need for this kind of photography often arises among students of all levels in various circles for the study of nature, in practical lessons in natural science, among workers'-faculty and university students, among scientific workers, and in laboratories — factory, analytical, scientific-research and others.

In the course of solving the task of microphotographing, a number of firms have put out installations, in principle identical but constructively various. Part of them, in general outline, may be reduced to the following: a massive plate A (left drawing) is provided with an upright B, strictly perpendicular to it; along it moves a washer C, connected with a frame D. The latter is provided with ground glass, a focal-plane shutter and a conical bellows E, the narrow end of which bears a device for fastening onto the eyepiece of the microscope.

The shot is made thus: the eyepiece of the microscope is replaced by a projection eyepiece, the microscope is set on the plate, and the focusing is done for clear vision. One sets the frame over the microscope so that its centre coincides with the optical axis of the microscope; after this the lower end of the bellows is fastened on the eyepiece, and by means of the micrometric screw of the microscope the needed sharpness of the image is attained on the ground glass, after which the cassette is set in place and the shot is made with the corresponding exposure. The degree of magnification of the object on the shot is the greater, the greater the extension of the bellows at the shot. The method of exactly determining the magnification will be described below.

The considerable cost of the described installation makes it little available to the broad circle of persons who need it. This circumstance compels a wide use of drawing. For greater accuracy of the latter, special devices are used, joined to the eyepiece and allowing one to see with one eye both the drawing and the object at the same time. This method too, however, is not accessible to all, since it requires the ability to draw.

Some time ago I took up the testing of simplified methods of microphotographing in general, and of the applicability for this of an ordinary photo-camera in particular, and established the full possibility of applying the latter — but only with ground glass.

The essence of the method is this. On the plate of an ordinary laboratory stand A (right drawing) is set, as in the previous case, the microscope, and the focusing is done for clear vision. After this, by means of the coupling B and the rod C, which has a thread corresponding to the nut of the camera, the camera E is fastened. Its lens is placed at a distance of 1–2 mm from the eyepiece and is centred with it. The ground glass is set parallel to the stage of the microscope, and by means of the micrometric screw the required sharpness is set on the ground glass. Then the shot is made. The magnitude of the exposure depends on the object, the lighting, and the diaphragming of the microscope lens, and varies from fractions of a second to several minutes.

Diaphragming the microscope lens gives a great depth of sharpness, and with lenses of low quality eliminates spherical aberration. Diaphragming the camera lens gives a weak effect in the same direction, and, with strong eyepieces, considerably reduces the field of the image, which in most cases is undesirable. In practice the diaphragming of the microscope lens is done by inserting into it little circles of dense black paper with an opening of the corresponding diameter in the centre.

In shooting non-transparent objects one has to use reflected light. At magnifications up to 200–250 times one may use side light; at greater magnifications it is necessary to illuminate the object through the lens, by means of a special mirror or prism placed in the lens or in the tube of the microscope.

The degree of magnification is determined by the extension of the camera bellows. At single extension the magnification of the image on the ground glass is half that which the microscope gives, in the given case, to the eye. At double extension the degree of magnification on the ground glass coincides with the magnification of the microscope. For other extensions the magnification obtained can be determined exactly by means of an objective micrometer. The latter is a slide with millimetre divisions marked on it; one of these divisions is divided further into 10 or 100 parts, and thus represents 0.1 or 0.01 mm respectively.

The measurement is made thus. The slide with the preparation is taken off the stage of the microscope, and in its place is set the micrometer. Then the needed sharpness of the scale on the ground glass is attained, and a measurement with a millimetre scale is made of one or another of its divisions. Dividing the figure obtained by the measured one, we get the exact magnitude of the magnification in the given case. For example: 1 mm of the micrometer scale proved equal on the ground glass to 100 mm — the magnification will be:

100 : 1 = 100

that is, 100-fold. Or another example: 0.1 mm of the micrometer scale equals 50 mm on the ground glass; the magnification, consequently, will be:

50 : 0.1 = 500

that is, 500-fold.

In the absence of a ready micrometer, one can make it sufficiently exactly oneself. For this one must shoot a millimetre scale so that the image comes out 10 or 100 times smaller; from the obtained negative a diapositive is made, and from it is cut a strip with the scale, which can serve as an objective micrometer.

On the attached shots is depicted my installation, built by the above-described scheme, and samples of the microphotographs obtained with its help. The upper shot — crystals of potassium nitre at a magnification of 60 times; their lighting is from the side. The lower shot was made under lighting by transmitted light and represents the hind legs of a flea at a magnification of 130 times. The extension of the camera bellows at the 1st shot is single, at the 2nd double.

The shots are made on orthochromatic plates (sensitivity 90 by Wynne), “Aerofoto.” The source of light was an opal half-watt lamp of 60 watts, placed at a distance of 20 cm from the mirror of the microscope. The diameter of the opening in the diaphragm of the microscope lens was 2 mm; the lens of the photo-camera was diaphragmed to F = 192. The exposure time was as follows: for the 1st shot — 2 minutes, and for the 2nd shot — 3 minutes 20 seconds.

Thus, the presence of a photo-camera with ground glass allows one to solve well enough the task of obtaining the needed microphotograph of this or that preparation.

Besides the described microphotographs at large magnifications, in a number of cases photographs with a magnification of 2 to 20 times are needed. This is easily attained by attaching to the camera lens a suitable magnifier and diaphragming to the desired sharpness. Such a device can be successfully applied in shooting snowflakes, small plants, insects, seeds and the like.

— Engineer-technician A. M. Belousov